BRONCHOSCOPE DISINFECTION GUIDE for OES BF 40 Series Bronchofiberscopes Refer to the companion BRONCHOSCOPE CLEANING GUIDE tor detailed cleaning information WARNING This guide is only a summary of the steps necessary to properly disinfect your bronchoscope Be sure to follow the detailed steps outlined in the ENDOSCOPE REPROCESSING MANUAL that was included with your bronchoscope when purchased HIGH LEVEL DISINFECTION 2 Fill basin with disinfectant Attach a suction cleaning adapter MAJ 222 to the previously cleaned bronchoscope Immerse completely in disinfectant solution Attach a syringe to the adapter Pull the plunger of the syringe to fill the instrument suction channels and channel cleaning adapter with disinfectant solution RINSING ...